The OX-TRAN 2/22 series of 1-cell analyzers can test oxygen transmission rate (OTR) on films and packages from low- to moderate-level barriers. They feature one changeable horizontal test cartridge.
InquiryThe 2/22 Series uses MOCON’s patented COULOX absolute (intrinsic) coulometric sensor requires no calibration and is the base for the ASTM standard D3985. It is the only instrument family proven to meet its precision and bias requirement through independent and 3rd-party testing.
ModelsThis family of highly accurate instruments has 2 models having a different lower detection limit.
Their upper detection limit of 200 cc/(m2 × day) can be extended with the use of the reduced area film test cartridges, atmospheric oxygen in place of the normal 100% O2 (factor of 4.8) or various size masking. The series has industry leading features such as removable and interchangeable horizontal test cartridges, easy to use touchscreen operator interface, and fully automated testing providing higher throughput in the lab.
For more than 25 years, MOCON’s OX-TRAN series of instruments have been the most trusted and longest lasting on the market. From simple single layered barriers to complex multi-layered barriers, the OX-TRAN 2/22 series is the right choice.
Removable Cartridges| Parameter | Units | Model H | Model A |
|---|---|---|---|
| Test Ranges (50 cm²) | cc/(m² × day) | 0.05 to 200 | 0.1 to 200 |
| Test Ranges (50 cm²) | cc/(100 in² × day) | 0.003 to 12.9 | 0.006 to 12.9 |
| Test Ranges (5.6 cm²) | cc/(m² × day) | 0.44 to 1770 | 0.88 to 1770 |
| Test Ranges (5.6 cm²) | cc/(100 in² × day) | 0.027 to 114.0 | 0.054 to 114.0 |
| Test Ranges (5.0 cm² masked) | cc/(m² × day) | 0.5 to 2000 | 1.0 to 2000 |
| Test Range - Package | cc/(pkg × day) | 0.00025 to 1.0 | 0.0005 to 1.0 |
| Resolution (50 cm²) | cc/(m² × day) | 0.0001 | 0.0001 |
| Repeatability | cc/(m² × day) | 0.02 or 1%* | 0.02 or 1%* |
| Specification | Units / Details |
|---|---|
| Test Temperature Range | 20-40 °C +/- 0.2 °C |
| Controlled RH Testing Range, Test Gas | 0%, 30-90% +/- 3% |
| Controlled RH Testing Range, Carrier Gas | 0%, 30-90% +/- 3% |
| Film Size | 4” x 4” Std, 2” x 2” optional |
| Film Thickness | Up to 120 Mil |
| Film Test Area | 50 cm² std., option to 5.6 cm² |
| Specification | Details |
|---|---|
| Dimensions | 15.5” H x 12” W x 23” D (39.4 cm x 30.4 cm x 58.0 cm) |
| Weight | 95 Lbs. (43.1 kg) |
| Specification | Details |
|---|---|
| Voltage | 100 - 240 VAC 50/60 Hz |
| Maximum Power Draw | 700 VA |
| Condition | Details |
|---|---|
| Temperature (Operation) | 22 °C ± 2 °C |
| Temperature (Storage) | 10 °C to 30 °C |
| Humidity (Operation) | 20% to 80% RH (non-condensing) |
| Humidity (Storage) | 5% to 85% RH (non-condensing) |
| Specification | Details |
|---|---|
| Carrier Gas Composition | Nitrogen/Hydrogen Mix (98%/2%) |
| Carrier Gas Supply Pressure | 25 PSI (1.72 Bar), (172 kPa) |
| Test Gas Composition | 100% Oxygen (purity 99.9%) |
| Test Gas Supply Pressure | 25 PSI (1.72 Bar), (172 kPa) |
| Specification | Details |
|---|---|
| Remote Access and Control | Requires optional PERM-NET Lite software |
| Ethernet | Standard Ethernet port (100T) |
| USB | Dual USB 2.0 ports for keyboard, mouse or memory devices |
| Specification | Details |
|---|---|
| Type of equipment | Electrical Equipment for measurement, Control and Laboratory Use |
| Analyzer Type | Permeation Analyzer |
| Machinery Directive | 2006/42/EU |
| EMC Directive | 2014/30/EU |
| RoHS Directive | 2011/65/EU |
| Standards | IEC/EN 60204-1:2018; IEC 61010-1:2010+A1 2019; EN 61326-1:2021 |

MOCON
Lower-cost Oxygen Transmission Rate (OTR) analyzer for films from low to high barriers.

MOCON
High-performance 2-cell OTR analyzer for films and packages with very high barriers.

MOCON
Four-cell OTR analyzer providing high-throughput barrier testing.

MOCON
Advanced whole-package OTR analyzer with integrated environmental chamber.

MOCON
Eight-cell OTR analyzer for ambient testing of complete packages.