Inquiry| Specification | LeakPointer 3 | LeakPointer 3+ |
|---|---|---|
| Applications | Non-destructive leak tests on individual flexible or rigid MAP packages | Non-destructive leak tests on individual or multiple flexible or rigid MAP packages |
| Vacuum Pressure | 50-750 mbar | 50-800 mbar |
| Dimensions (lid open) W x D x H | 400 x 617 x 698 mm | 555 x 812 x 981 mm |
| Weight | 20 kg | 50 kg |
| Power Supply and Consumption | 100-264 VAC + PE, 50/60 Hz; max. 50W | (115V) 103-127 VAC + PE, 57-63 Hz; max. 900W / (230V) 207-253 VAC + PE, 47-53 Hz; max. 900W |
| Configuration | W1 | W2 | H1 | H2 |
|---|---|---|---|---|
| LeakPointer 3 With PackFix | 325 mm | 295 mm | 40 mm | 86 mm |
| LeakPointer 3 Without PackFix | 325 mm | 268 mm | 40 mm | 110 mm |
| LeakPointer 3+ With PackFix | 465 mm | 420 mm | 40 mm | 117 mm |
| LeakPointer 3+ Without PackFix | 465 mm | 363 mm | 40 mm | 155 mm |
| Specification | Details |
|---|---|
| Air Supply and Consumption | 5.5 ± 0.5 bar (80 ±7 psi); Ø6/4 mm hose / Approx. 33 L per measurement; max. 162 L/min |
| Integrated Vacuum Pump | Included |
| Sensor Type | NDIR CO₂ sensor, single beam |
| Test Time | 10-35 seconds |
| Measurement Units and Accuracy | Single package mode: estimated diameter of reference hole in microns (minimum 50 microns) / Multiple package mode: measured increase of CO₂ in ppm/s |
| CO₂ Content in Packages | Minimum 10% |
| Ambient CO₂ Level | Max 4500 ppm, recommended <1500 ppm |
| Ambient Temperature | Operational: +2°C to +35°C / Storage: -20°C to +60°C |
| Relative Humidity | Operational: +2°C to +25°C: 10 to 90 %RH / +25°C to +30°C: 10 to 70 %RH / +30°C to +35°C: 10 to 50 %RH (non-condensing) / Storage: <95 %RH |
| Number of Test Definitions | 100 |
| Data Collection Capacity and Export | Stores test data from up to 1,000,000 tests; export to CSV file |
| Connectivity | LAN: RJ-45 Ethernet 10/100 Mbit/s, DHCP client or fixed IP / USB: 2 x USB 2.0 Type A |
| Accessories | PackBase (1 included for units sold with PackFix), PackFix, PackTray, Leak Simulation Kit, Barcode scanner |
| Calibration and Service | 12 months |
| Compliance | CE, China RoHS |

Dansensor
High-accuracy tabletop headspace MAP gas analyzer.

Dansensor
Portable headspace MAP gas analyzer for packaging lines.

Dansensor
Multi-function tester for seal strength, burst, creep, and leak tests.

Dansensor
Vacuum leak detector for blister packs, vials, and rigid packages.

Dansensor
In-line leak detector to reduce waste and product returns.